Fault

Possible cause

Remedy

❑ Displayed pressure value

is fluctuating.

➨ The MPT 100 / MPT 200 was

exposed to a significant pres-

sure change?

✔ Allow for a stabilization time

of 5-10 minutes.

➨ The vacuum gauge head

MPT 100 / MPT 200 is still in

the stabilization phase?

✔ Allow for a stabilization time

of 5-10 minutes.

➨ Pressure measurement at a

pressure above 50 mbar?

✔ Deviation within the range of

the measurement uncertain-

ty. Use pressure transducer

VSK 3000 (see accessories).