Fault
Possible cause
Remedy
❑ Displayed pressure value
is fluctuating.
➨ The MPT 100 / MPT 200 was
exposed to a significant pres-
sure change?
✔ Allow for a stabilization time
of 5-10 minutes.
➨ The vacuum gauge head
MPT 100 / MPT 200 is still in
the stabilization phase?
✔ Allow for a stabilization time
of 5-10 minutes.
➨ Pressure measurement at a
pressure above 50 mbar?
✔ Deviation within the range of
the measurement uncertain-
ty. Use pressure transducer
VSK 3000 (see accessories).