Measuring Power or Loss

OTDR

85

Referencing Your Power Meter to a Source

4. Activate the source at the desired wavelength.

5. Match the power meter wavelength with the

source wavelength.

Press Wavelength (F1/F2 key) to switch
between

favorite wavelengths

of your power

meter (see Defining the List of Favorite
Wavelengths
on page 80).

6. Press W/dBm/dB (F1/F2 key) until you get dB units to retrieve the last

saved reference.

OR

Press Reference (F1/F2 key) to save the
current power as the new reference.

Reference power appears (in dBm) and
current loss is automatically switched to dB.

7. Repeat the procedure for each wavelength you want to reference.