Hanna Instruments HI 83099 User Manual

Page 45

Advertising
background image

89

88

Nitrite HR

• Add the content of one packet of HI 93708-0 reagent.

Replace the cap and shake gently until completely
dissolved.

• Reinsert the cuvette into the instrument.

• Press

Timer and the display will show the countdown prior to the measurement or, alternatively, wait

for 10 minutes and press

Read. When the timer ends the meter will perform the reading. The

instrument displays concentration in

mg/L of nitrite.

• Press ▲ or ▼ to access the second level functions.
• Press the

Chem Frm key to convert the result in mg/L of nitrogen-nitrite (NO

2

¯-N) and sodium nitrite

(NaNO

2

).

• Press ▲ or ▼ to return to the measurement screen.

Nitrite HR

NITRITE HIGH RANGE

SPECIFICATIONS

Range

0 to 150 mg/L

Resolution

1 mg/L

Accuracy

±4 mg/L ±4% of reading at 25 °C

Typical EMC

±1 mg/L

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 575 nm

Method

Adaptation of the Ferrous Sulfate method. The reaction between nitrite and the
reagent causes a greenish-brown tint in the sample.

REQUIRED REAGENTS

Code

Description

Quantity

HI 93708-0

Powder reagent

1 packet

REAGENT SETS

HI 93708-01 Reagents for 100 tests
HI 93708-03 Reagents for 300 tests
For other accessories see page 132.

MEASUREMENT PROCEDURE

• Select the

Nitrite HR method using the procedure described in the

Method Selection section (see page 12).

• Fill the cuvette up to the mark with 10 mL of unreacted sample and

replace the cap.

• Place the cuvette into the holder and close the lid.

• Press the

Zero key. The display will show “-0.0-” when the meter is zeroed and ready for

measurement.

• Remove the cuvette.

10 mL