5 Battery Information

English

9

How to Operate and Adjust Suction level

1

Unwind the Patient Tube (check that no kinks will obstruct

flow).

2

Turn the unit “ON” by pressing the button ( ).

3

Block the Patient Tube.

4

3

4

Set desired Suction Vacuum level by

turning the Vacuum Regulator.

• Turn clockwise (+) to increase vacuum

• Turn counter-clockwise to decrease

vacuum

5

The Vacuum level will display on the 50 – 550 mmHg scale.

6

When the desired Vacuum level is reached; unblock and

then block the Patient Tube.

Result: Unit should return to the desired level.

7

Apply necessary suction therapy. If desired, a suction tip or

catheter could be attached to the Patient Tube.

Important
If LCSU 4 does not maintain the desired suction

performance level, refer to Troubleshooting instructions

(see Chapter 7).

After Each Use

1

When finished with suctioning, let LCSU 4 run for a

moment to allow all suctioned material to flow from the

Patient Tube into the Canister.

2

Disconnect and dispose of Canister and Patient Tube.

3

Clean the exterior and any reusable parts of the LCSU 4

assembly according to instructions (see Chapter 6).

4

Perform Device Test (see Chapter 6).

5

Place Battery on charge (see Chapter 5).

Battery Charging

Important
Use only Laerdal Battery Cat. No. 886113.

Caution
Use of other than Laerdal brand Battery may result in

errors related to the Battery status indicator, reduction of

the Battery operation time, failure to effectively operate

LCSU 4, and/or give rise to hazards to operator and/or

patient.

When in operational use
1

An empty Battery must charge up to 5 hours to reach full

capacity.

2

Battery run time: Approximately 45 minutes of continuous

operation at zero vacuum level (free flow), fully recharged.

3

Always fully charge Battery - before and after each use.

To prolong Battery lifetime it is recommended to place the

Battery on continuous charge. It will not harm the unit.

If continuous charging is not possible, charge Battery for

minimum 24 hours once a month. Completely discharging the

Battery will shorten its useful life.

When stored unused for > 3 months
• Fully charge Battery before storing
• Recharge every 3-6 months

Battery Test

When in operational use, the Battery should be tested every

6-12 months.

1

Start the test with a fully charged Battery.

2

Set Max Vacuum level.

3

Let the unit run for 20 minutes (free flow).

4

Block the Patient Tube.

5

If Vacuum level fails to reach 550 mmHg the Battery

should be replaced.