2 technical data, 2technical data – Nor-Cal CDG 025 OP Lit User Manual

Page 9

Advertising
background image

CDG025-OP-LIT (2010-1)

9

2

Technical Data

Measurement range

→ "Validity"

Accuracy

1

0.20% of reading

)

Temperature effect on zero

≥10 Torr (F.S.)
1 / 2 Torr (F.S.)


0.0050% F.S./ °C
0.015% F.S./ °C

Temperature effect on span

0.01% of reading / °C

Resolution

0.003% F.S.

Gas type dependence

none

Output signal analog
(measuring signal)

Voltage range

–5 … +10.24 V

Measuring range

0 … +10 V

Relationship voltage-pressure

linear

Output impedance

0

Ω (short-circuit proof)

Loaded impedance

>10 k

Response time

30 ms

Gauge identification

Resistance 13.2 k

refer-

enced to supply common
(Voltage at pin 10

≤5 V)

RS232C interface

Transmission rate
Data format




Connection


9600 baud
binary
8 data bits
one stop bit
no parity bit
no handshake
→ "Electrical Connection"

Further information about the RS232C interface

 [1]).

1)

Non-linearity, hysteresis, repeatability in the calibrated range at 25 °C

ambient operating temperature without temperature effects after
operation of 2 h.

Advertising