ROHO SelectAir MA mattress User Manual
Page 35
33
Section 14
HybridSelect
To deactivate the Upright Mode, first lower the patient to a lying position.
Then press the "Upright On/Off" button a second time (Ensure the upright
light extinguishes).
Note: Do not leave the control unit in the Upright Mode when the
patient is lying flat or at a slight angle.
Note: Ensure the patient has at least one (1) inch (3 cm) of clearance
between the bed frame and their body. If not, manually increase or
decrease system pressure to compensate.
Firm Mattress Mode
To aid in the placement and maneuvering of the patient, the control unit is
equipped with a Firm Mattress Mode. The Firm Mattress Mode will fully inflate
the mattress making it firm for easy positioning. The Firm Mattress Mode will
automatically terminate when the time displayed on the keypad has expired.
Press the "Firm" button on the keypad. The display will indicate the
remaining time (fig. 36).
To end the cycle press the Firm Mattress “Off” button.
Note:
The HybridSelect control unit is equipped with an internal timer which
will turn the Firm Mattress Mode off after 25 minutes. If the Firm
Mattress Mode has timed out and ended the firm mattress cycle, DO
NOT press the Firm "Off” key as this may cause the system to "lock up".
Note:
Pressing the Firm "Off" key multiple times or holding down the button
may cause the system to “lock up”.
Note:
To ensure patient comfort, the control unit slowly lowers the pressure of
the system. This may give the appearance that the firm function is still
active. If the timer lights are extinguished, the control unit is in its
normal mode.
System Lock
Control unit is equipped with the System Lock which suspends normal keypad
operations (with exception of Firm Mattress, Upright and CPR on/off) until deactivated.
When the Firm Mattress button is pressed while the System Lock is activated, the
System Lock light will extinguish until the Firm Mattress Mode turns off.
1. To lock the control panel, press the “System Lock” button (fig. 37), the System
Lock light will come on.
2.
1.
2.
Light
Figure 37
Figure 36