Appendix G: MANUALLY INITIALIZING SENSORS

Manually Initializing the Carbon Monoxide and Oxygen Sensors

If the CO or O2 sensor fails to initialize, this procedure can be performed to
try and manually initialize the sensor(s).

WARNING : Changing parameters in field calibration mode can adversely
affect the operation of your analyzer. Follow the instructions carefully.

1. Press and hold the Scroll/Enter Key down until ‘FILD CAL’ is displayed. The

714 will cycle through a 30 second countdown then display ‘CAL’, Air’, gAS’.

2. Use the Up Arrow Key to select ‘Air’ by making it blink. Press the Scroll/Enter

Key to accept the selection.

3. 714 Display: ‘CAL’, ‘FLo’, ‘oPEn’. Action : Press the Func Key to bypass.

4. 714 Display: ‘CAL’, ‘FLo’, ‘CLos’. Action : Press the Func Key to bypass.

5. 714 Display: ‘0ppm’, ‘Co’, ‘rEF’. Action : Press the Scroll/Enter Key. After the

10 second countdown is complete press the Scroll/Enter Key twice.

6. 714 Display: ‘20.9%’, ‘o2’, ‘rEF’. Action : Press the Scroll/Enter Key. After the

10 second countdown is complete press the Scroll/Enter Key twice.

7. 714 Display: ‘0ppm’, ‘No’, ‘rEF’. Action : Press the Scroll/Enter Key. After the

10 second countdown is complete press the Scroll/Enter Key twice.

8. 714 Display: ‘0.0mbar’, ‘0.0’, ‘oFS’. Action : Press the Func Key to bypass.

9. 714 Display: ‘2000’, ‘A-r’, ‘SEt’. Action : Press the Func Key to bypass.

10. 714 Display: ‘CAL’, ‘StoP’, ‘go’. Action : Press the Up Arrow Key to select

‘StoP’ by making it blink and press the Scroll/Enter Key to accept the selec-
tion. The 714 will return to normal operation

If the 714 continues to fail to automatically initialize or this procedure fails
to initialize the sensors, return your analyzer to Test Products
International.

Appendix F: CO ALARM SETTINGS

Setting the Carbon Monoxide Alarm Point and/or Auto Power Off Mode

The CO alarm and auto power off settings can be changed as necessary by per-
forming the following procedure.

WARNING : Changing parameters in field calibration mode can adversely affect
the operation of your analyzer. Follow the instructions carefully.

1. Press and hold the Scroll/Enter Key down until ‘FILD CAL’ is displayed. The 714

will cycle through a 30 second countdown then display ‘CAL’, Air’, gAS’.

2. Use the Up Arrow Key to select ‘Air’ by making it blink. Press the Scroll/Enter

Key to accept the selection.

3. 714 Display: ‘CAL’, ‘FLo’, ‘oPEn’. Action : Press the Func Key to bypass.

4. 714 Display: ‘CAL’, ‘FLo’, ‘CLos’. Action : Press the Func Key to bypass.

5. 714 Display: ‘CAL’, ‘LEAK’, ‘rEF’. Action : Press the Func Key to bypass.

6. 714 Display: ‘0ppm’, ‘Co’, ‘rEF’. Action : Press the Func Key to bypass.

7. 714 Display: ‘20.9%’, ‘o2’, ‘rEF’. Action : Press the Func Key to bypass.

8. 714 Display: ‘0.0mbar’, ‘0.0’, ‘oFS’. Action : Press the Func Key to bypass.

9. 714 Display: ‘2000’, ‘A-r’, ‘SEt’. Action : If you are not changing the CO alarm

point press the Func Key. Otherwise press the Up and Down Keys to set the
ppm level the alarm will sound at. Factory default is 2000ppm. Press the
Scroll/Enter Key to accept the selection.

10. 714 Display: ‘CAL’, ‘StoP’, ‘go’. Action : Press the Up Arrow Key to select

‘StoP’ by making it blink and press the Scroll/Enter Key to accept the selection.
The 714 will return to normal operation