Monroe Electronics Charge Plate - Ionizer Performance Analyzer - model 287A User Manual
Page 7
iii) The test results are displayed at the end of the test. The results are
displayed in three windows which may be scrolled with either the or
button. The first window shows the positive and negative decay times
(from 1000 volts to 100 volts) for tests number 1 and 2 in the sequence.
The second window shows decay times for test number 3 and the
average times for three + and three – decay tests. The third window
shows the + and – balance peak voltages and the temperature and
humidity taken during the test sequence.
c)
Press ‘ENTER’ to return to the main menu.
d)
Repeat a, b and c to gather a second set of data. This will be ‘Test 2’.
5.
Run manual decay tests —
a)
Press and release ‘+ CHARGE’ and let the first manual decay test run its
course and display a result. That manual decay test will be labeled ‘M1+’. To
return to the main menu, press ‘ENTER’.
b)
Press and release ‘+CHARGE’ again and run the test to completion. This
second manual decay test will be labeled ‘M2+’. To abort one of these tests at
any time during the test press the button that got the test started.
c)
Return to the main menu. Press and release ‘
CHARGE’. This third manual
decay test will be labeled ‘M3
’ when completed. Return to the main menu.
6.
Run manual balance tests —
a)
Press and hold either the or button until Balance test is indicated on the
display then release. This will start your fourth manual test. It will be labeled
‘M4B’ when finished. Return to the main menu.
b)
Repeat for a fifth manual test. This test will be labeled ‘M5B’. Note that if you
hold the button down too long to start the test, the test will be nullified.
Following test ‘M5B’, return to the main menu. The memory usage will be
displayed as ‘A20%M50%’. You have conducted 2 out of 10 automatic
sequences and 5 out of 10 manual tests.
7.
At this point, you may elect to shut the equipment down and retire to a different
location to contemplate the results.
Test results —
Press and release the ‘ON-OFF’ button. You will be prompted for a Y/N? response to the
ClrData (clear data) question. NO!! Press ‘ENTER’ to accept “N”. That will take you to the
main menu while retaining the data.
5