25

(2) Current

* The measured value <1000dgt, add 5 dgt to the accuracy.
Overload protection :
600A

rms


Position Error :
±1% of reading.
AC Conversion Type and additional accuracy is same as AC
Voltage.


AC+DC Arms Accuracy : Same as ACA spec + DCA spec.

DCA affected by the temperature and the residual magnetism.

Press HOLD key > 2sec to compensate it.

Function

Range

Accuracy

DCA

99.99A

± (1.5% + 0.2A)

599.9A

± (1.5% + 5dgt)

*

ACA

0.10A ~ 99.99A

50 ~ 60Hz ± (1.5% + 5dgt)

*

>60 ~ 400Hz ± (2% + 5dgt)

*

599.9A

LPF

ACA

0.10A ~ 99.99A

50 ~ 60Hz ± (1.5% + 5dgt) *

>60 ~ 400Hz ± (5% + 5dgt)

*

599.9A

(3) Peak Hold : Peak MAX / Peak MIN

ACV

140.0V

± (3.0% + 15dgt)

1400V

ACA

140.0A

± (3.0% + 15dgt)

850A

Function

Range

Accuracy

Overload protection : 1000 V

rms

/ 600 A

rms

Accuracy defined for :
Sine wave, ACV>5Vrms / ACA≧5Arms, Freq.50~400Hz.

Only suitable for the repetitive events.