Pyramid probe core off-line cleaning with a brush, Step 4: blow away particles and solvent – Cascade Microtech Pyramid Probes User Manual

Page 3

Advertising
background image

PN 153-689-1

www.cascademicrotech.com

Off-line Core Cleaning With a Brush• 3

Step 4: Blow Away Particles and Solvent

Complete the following steps to blow the remaining IPA or methanol containing particles/residue away from the probe face.
1. Hold an acceptable air nozzle vertical to the probe face, three inches (75 mm) from the core (see

Figure 4

). Activate the

air stream before approaching the core with the air nozzle, as the nozzle will tend to move forward even at low pressure.

2. Continue blowing air on the probe face until it is visibly dry, approximately 30-60 seconds.
3. Return to

Step 1: Inspecting the Core

, and repeat steps 2, 3 and 4 until all particles have been removed.

Corporate Headquar

Cascade Microtech, Inc.

Germany

phone:

+49-89-9090195-0

email:

cmg_sales

@cmicro.com

Japan

phone:

+81-3-5615-5150

email:

cmj_sales

@cmicro.com

China

phone:

+86-21-3330-3188

email:

cmc_sales

@cmicro.com

e

Singapor

phone:

email:

cms_sales

@cmicro.com

Taiwan

phone:

+886-3-5722810

email:

cmt_sales

@cmicro.com

toll free:

+1-800-550-3279

phone:

+1-503-601-1000

email:

cmi_sales

@cmicro.com

+65-6873-7482

ters

© Copyright 2011 Cascade Microtech, Inc. All
rights reserved. Cascade Microtech is a
registered trademark of Cascade Microtech,
Inc. Pyramid Probe is a trademark of Cascade
Microtech, Inc. All other trademarks are the
property of their respective owners.

Data subject to change without notice.

Pyramid Probe Core Off-line Cleaning With a Brush

Advertising