Pyramid probe core off-line cleaning with a brush, Step 4: blow away particles and solvent – Cascade Microtech Pyramid Probes User Manual
Page 3
PN 153-689-1
www.cascademicrotech.com
Off-line Core Cleaning With a Brush• 3
Step 4: Blow Away Particles and Solvent
Complete the following steps to blow the remaining IPA or methanol containing particles/residue away from the probe face.
1. Hold an acceptable air nozzle vertical to the probe face, three inches (75 mm) from the core (see
Figure 4
). Activate the
air stream before approaching the core with the air nozzle, as the nozzle will tend to move forward even at low pressure.
2. Continue blowing air on the probe face until it is visibly dry, approximately 30-60 seconds.
3. Return to
, and repeat steps 2, 3 and 4 until all particles have been removed.
Corporate Headquar
Cascade Microtech, Inc.
Germany
phone:
+49-89-9090195-0
email:
cmg_sales
@cmicro.com
Japan
phone:
+81-3-5615-5150
email:
cmj_sales
@cmicro.com
China
phone:
+86-21-3330-3188
email:
cmc_sales
@cmicro.com
e
Singapor
phone:
email:
cms_sales
@cmicro.com
Taiwan
phone:
+886-3-5722810
email:
cmt_sales
@cmicro.com
toll free:
+1-800-550-3279
phone:
+1-503-601-1000
email:
cmi_sales
@cmicro.com
+65-6873-7482
ters
© Copyright 2011 Cascade Microtech, Inc. All
rights reserved. Cascade Microtech is a
registered trademark of Cascade Microtech,
Inc. Pyramid Probe is a trademark of Cascade
Microtech, Inc. All other trademarks are the
property of their respective owners.
Data subject to change without notice.
Pyramid Probe Core Off-line Cleaning With a Brush