Ocx 8800, Instruction manual, Specifications – Emerson ROSEMOUNT OCX 8800 User Manual
Page 17
Instruction Manual
IM-106-880, Rev 1.0
January 2007
1-9
OCX 8800
SPECIFICATIONS
Hazardous Area OCX
Specifications
Net O
2
Range
0-1% to 0-40% O
2
, fully field selectable
Combustibles
0-1000 ppm to 0-5%, fully field selectable
Accuracy
Oxygen
± 0.75% of reading or 0.05% O
2
(whichever is greater)
Combustibles
± 2% range
System Response to
Test Gas
Oxygen
10 sec T90
Combustibles
25 sec T90
Temperature Limits
Process
32° to 2600°F (0° to 1427°C)
Sensors Housing
-40° to 212°F (-40° to 100°C), ambient
Electronics Housing
-40° to 149°F (-40° to 65°C), ambient
-40° to 185°F (-40° to 85°C), internal - operating temperature of
electronics inside instrument housing, as measured by a HART
communicator or AMS software
Nominal and Approximate
Shipping Weights
18 in. (457 mm)
probe package
54 lbs (20 kg)
3 ft (0.91 m) probe
package
55 lbs (20.5 kg)
6 ft (1.83 m) probe
package
57 lbs (21 kg)
9 ft (2.74 m) probe
package
59 lbs (22 kg)
Mounting and Mounting
Positions
Sensors Housing
Flange
Electronics Housing
Wall/Pipe
Materials
Probes
316L stainless steel - 1300°F (704°C)
Inconel 600 - 1832°F (1000°C)
Ceramic - 2600°F (1427°C)
Enclosures
Low-copper aluminum
Calibration
Semi-automatic or automatic
Calibration Gas Mixtures
Recommended
(Ref. test gas bottles
kit #1A99119G04)
0.4% O
2
, Balance N2
8% O
2
, Balance N2
1000 ppm CO, Balance Air
Calibration Gas Flow
7 scfh (3.3 l/m)
Reference Air
2 scfh (1 l/m), clean, dry instrument-quality air (20.95% O
2
),
regulated to 45 psi (310 kPa)
Eductor Air
5 scfh (2.5 l/m), clean, dry, instrument-quality air 20.95% O
2
),
regulated to 45 psi (310 kPa)
Dilution Air
0.1 scfh (0.5 l/m), clean, dry, instrument-quality air (20.95% O
2
)
regulated to 45 psi (310 kPa)
Blowback Air (optional)
Clean, dry, instrument-quality air (20.95% O
2
), regulated to 55 psi
(379 kPa)
Table continued on next page