Specifications:
Dimensions:
Height: 10.1 in (25.7 cm)
Width: 6.5 in (16.5 cm)
Depth: 4.5 in (11.4 cm)
Weight:
6.8 lbs (3.08 kg)
Shipping Weight:
13.1 lbs (5.94 kg)
Altitude:
Sea level to 9000 ft (2743 m)
Operating Temperature:
41˚F to 104˚F (5˚C to 40˚C)
Storage / Transport Temperature: -4˚F to 140˚F (-20˚C to 60˚C)
Operating / Storage / Transport
Relative Humidity:
Up to 95% Non-Condensing
Oxygen Concentration:
87% to 95%
Start Up Time
87% O2 Concentration ≤ 10 min
Oxygen Minute Volume*:
Setting 1:
240 +/-15% mL/min
Setting 2:
380 +/-15% mL/min
Setting 3:
520 +/-15% mL/min
Setting 4:
660 +/-15% mL/min
Setting 5:
780 +/-15% mL/min
Mode of Delivery:
Pulsed Dose only
Trigger Sensitivity:
0.50 cmH2O maximum
Trigger Delay:
0.11 seconds typical
Power:
AC Power Adapter: 100-240 VAC (1.0 A @ 120 VAC)
DC Power Adapter: 11-16 VDC (5.0 A @ 12 VDC)
Battery Duration:
Setting 1: 4.5 hrs
(Approximate)
Setting 2: 3.2 hrs
Setting 3: 2.5 hrs
Setting 4: 1.8 hrs
Setting 5: 1.5 hrs
Sound (@ Setting 2):
44 dBA
Cannula:
Maximum 7 ft (2.1 m) long standard adult single lumen nasal cannula.
Equipment Classification:
Class II - with respect to protection from electrical shock
Type BF - degree of protection against electrical shock
IPX1 - degree of protection against ingress of liquids
Mode of Operation - Continuous
Volatile Organic Compound (VOC) and Particulate Requirements:
The oxygen delivered from the EasyPulse POC meets the following requirements for particulate levels, VOC levels,
carbon monoxide levels, carbon dioxide levels and ozone levels.
EPA PM2.5
Particulate Matter
ASTM D5466
Ozone Levels
21 CFR 801.415
Carbon Monoxide Levels
EPA NAAQS -
Carbon Monoxide Carbon Dioxide Levels
OSHA Permissible Exposure Limits
Standard Test Method for Determination of Volatile Organic Chemicals in Atmospheres
(Canister Sampling Methodology)
* At Sea level