Alarm / indicators / troubleshooting, Warning – Precision Medical PM4351 EasyFlow5 User Manual

Page 9

Advertising
background image

8

7. Place the cannula over your

ears and position the prongs

in your nose as instructed

by your Provider or cannula

manufacturer.

OR

8. When the Concentrator is not in use, press the power On/Off ( / ) switch to

“OFF” ( ) position.

9. Unplug Concentrator from electrical outlet for complete disconnection from

Power Source.

ALARM / INDICATORS / TROUBLESHOOTING

The Audible Alarm and LED Indicators are listed in the chart below.

Visual Indicator/

Problem

Audible

Alarm

Probable Cause

Remedy

YELLOW indicator

flashes

Repeated

beeping

Flow to patient is

occluded

Remove occlusion, kink or block

in patient cannula

When device is

turned “ON”, device

does not function

and the GREEN

Power Indicator

does not light.

Continuous Concentrator is not

properly plugged into

electrical outlet.

Make sure Concentrator is

properly plugged into the electrical

outlet.

RED General

Alarm Indicator

is “ON” steady

or continuously

blinking; device

stops functioning.

Repeated

beeping

System malfunction.

Turn power switch “OFF” and

wait 5 minutes, then turn back

“ON”. If condition persists, turn

the concentrator “OFF”, connect

to another source of oxygen and

contact your Provider.

RED General Alarm

Indicator flashes

once, device

momentarily stops

functioning, then

automatically

restarts.

Single beep Momentary AC power

interruption.

It is normal for this to occur

occasionally (such as during

severe weather); however, if this

condition persists, connect to

another source of oxygen and

contact your Provider.

Continued on next page.

WARNING

ISO 8359 (Standard for Oxygen Concentrators) highly recommends that the

patient cannula that delivers gas to the patient from the oxygen concentrator

should include a Fire Stop Check Valve to stop the flow of gas towards the

patient in the case that the cannula becomes ignited. The Fire Stop Check

Valve should be located as close to the patient as reasonably practicable.

Fire Stop Check Valve - Precision Medical Part # 507706