Specifications – Precision Medical PM4351 EasyFlow5 User Manual

Page 7

Advertising
background image

5

Specifications:

Weights:

Device: 29.6 lbs (13.4 kg)

Device Boxed for shipping: 33.9 lbs (15.38 kg)

Dimensions:

Height: 17.0 in (43.2 cm)

Width: 12.25 in (31.1 cm)

Depth: 17.125 in (43.5 cm)

Power:

115 ±10% VAC; 60Hz 3.2 A; 350 watts

Oxygen Concentration:

≥87% - up to 6,900 Ft (2103.1 m)

≥85% - >6,900 Ft (2103.4 to 3048 m)

Outlet Pressure:

8 psi ±10%

Outlet Temperature:

Within 1.3 F of ambient

Sound Level:

53 dBA

Operating Conditions:

Temperature: 50°F to 95°F(10°C to 35°C)

Altitude: Up to 10,000 ft (3,048 m)

Storage Conditions:

Temperature: - 29.2°F to 140°F(- 34°C to 60°C)

Max. Humidity: 95% non-condensing

Cannula Requirement:

Maximum 7 foot long adult standard single lumen oxygen nasal cannula with a
maximum 40 ft crushproof extension tubing

Flow Settings (l/min):

0.5, 1.0, 1.5, 2.0, 2.5, 3.0, 3.5, 4.0, 4.5, 5.0

Flow Accuracy:

.5 l/min -0.0 / +400 ml of setting - up to 10,000 Ft (3048 m)

1.0 to 1.5 l/min ± 200 ml of setting - up to 6,900 Ft (2103.1 m)

±20% of setting 6901 to 10,000 Ft (2103.4 to 3048 m)

2.0 to 5.0 l/min ±10% of setting - up to 6,900 Ft (2103.1 m)

±20% of setting 6901 to 10,000 Ft (2103.4 to 3048 m)

Equipment Classification:

Class II - with respect to protection from electrical shock
Type BF - degree of protection against electrical shock
IPX1 - degree of protection against ingress of liquids
Mode of Operation - Continuous

Volatile Organic Compound (VOC) and Particulate Requirements:

The oxygen delivered from the EasyFlow Concentrator meets the following requirements for particulate levels,
VOC levels, carbon monoxide levels, carbon dioxide levels and ozone levels.
EPA PM2.5

Particulate Matter

ASTM D5466

Ozone Levels

21 CFR 801.415

Carbon Monoxide Levels

EPA NAAQS

Carbon Monoxide Carbon Dioxide Levels

OSHA Permissible Exposure Limits
Standard Test Method for Determination of Volatile Organic Chemicals in Atmospheres

Advertising