Table b-4. factory default calibration passwords, B.100 system:password:cdisable command, B.101 system:password:new command – KEPCO KLP Series Developer's Guide, Rev 1 User Manual

Page 129: B.102 system:password[:cenable]:state? query, B.103 system:security:immediate command, B.104 system:set command, B.100 syst, B.101 syst, B.102 syst, B.103 syst

Advertising
background image

KLP 031113

B-23

B.100 SYSTem:PASSword:CDISable COMMAND

SYST:PASS:CDIS

Syntax:

Short Form: SYST:PASS:CDIS <val>

Long Form: SYSTem:PASSword:CDISable <val>

Description: Clears the password enable state if the value matches the current password.

B.101 SYSTem:PASSword:NEW COMMAND

SYST:PASS:NEW

Syntax:

Short Form: SYST:PASS:NEW <old password>,<new password>
Long Form: SYSTem:PASSword:NEW <old password>,<new password>

Description: Establishes new password. The old (current) password is checked, then replaced by the new pass-

word.

B.102 SYSTem:PASSword[:CENable]:STATe? QUERY

SYST:PASS:STAT?

Syntax:

Short Form: SYST:PASS[:CEN]:STAT?

Long Form: SYSTem:PASSword[:CENable]:STATe?

Return Value: <int_value> 0 or 1 (1 = enabled, 0 = disabled)

Description: Returns a 1 if the password state is enabled or a 0 if it is disabled.

B.103 SYSTem:SECurity:IMMediate COMMAND

SYST:SEC:IMM

Syntax:

Short Form: SYST:SEC:IMM

Long Form: SYSTem:SECurity:IMMediate

Description: Initializes all NVRAM variable to factory defaults. This includes clearing all *SAV/*RCL locations,

returning all relay functions to factory default settings, and setting Quick Boot feature to OFF. This
command does not affect virtual model and corresponding protection limits, last setting recall register
or present operating conditions. NOTE: System password must be enabled prior to issuing this com-
mand (see PAR. B.99).

B.104 SYSTem:SET COMMAND

SYST:SET

Syntax:

Short Form: SYSTem:SET {CM0 | DC0 | DC1 | LF0 | LF1 | KL0 | KL1 | L100 | M40 | SCP}
Long Form: SYSTem:SET {CM0 | DC0 | DC1 | LF0 | LF1 | KL0 | KL1 | L100 | M40 | SCP}

Description: Establishes Device Clear, Line Feed, and Reset functions. Sending SYST:SEC:IMM sets LF1,

DC0, and RO0 (as if CM0 was sent). Sending SYST:SET CM1 sets the KLP to operate in compatible
mode and have all GPIB functions compatible with software version 2.9 and lower units. Sending
SYST:SET CM0 sets the unit to be fully SCPI 1997 compliant.

CM0

Establishes DC0, LF0, conditions described below (SCPI 1997 Standard compliance).

DC0

Device Clear functions per IEEE 488.2 Standard. (No effect on the device (power supply),
only clears internal status registers.)

LF0

Line Feed not provided upon empty buffer condition (no effect when access is via LAN
interface).

DC1

Device Clear functions identical to *RST (Output set to 0V, voltage mode and output set to
OFF except if RO1 (see below) is set.)

LF1

Line Feed provided if buffer is empty and a read is performed. (no effect when access is
via LAN interface).

KL0

Enables front panel controls.

KL1

Disables front panel controls.

L100

Number (100) of locations programmable using LIST commands.

M40

Number (40) of locations programmable using *SAV and *RCL commands.

SCP

SCPI Programming Language.

TABLE B-4. FACTORY DEFAULT CALIBRATION PASSWORDS

MODEL (** -1200 or -1.2K)

PASSWORD

KLP 10-150-**

1015

KLP 20-120-**

2012

KLP 36-60-**

3660

KLP 75-33-**

7533

KLP 150-16-**

1516

KLP 300-8-**

3008

KLP 600-4-**

6004

Advertising