HP Refrigerated Recirculating Chillers User Manual
Page 5
Notes: Refer to the serial number plate on rear of the chiller for model and electrical data.
Cooling Capacity, Watts x 3.41 = BTUs/hr. Performance specifications determined at ambient temperature of 20°C/68°F.
For 50Hz models, derate cooling capacity and pump flow by 17%.
External Pressure Reducing Assembly (for Positive Displacement Pump models) Steps down high outlet pressure to 10 to 45 psi. Cat. No 060302
5
6000 Series Standard Chillers with Magnetic Drive Centrifugal Pump
Model
6205
6305
6505
Operating Temperature
-5° to 40°C
-5° to 40°C
-10° to 40°C
Compressor
1/4 hp
1/3 hp
1/2 hp
Cooling Capacity @ 20°C
750 Watts
1200 Watts
1320 Watts
10°C
450 Watts
825 Watts
1055 Watts
0°C
100 Watts
375 Watts
550 Watts
Pressure at 0 Flow Rate
10 psi
10 psi
10 psi
Flow Rate @ 0 psi
4.3 gpm / 16.3 lpm
4.3 gpm / 16.3 lpm
4.3 gpm / 16.3 lpm
Reservoir Capacity
1.65 gal / 6.25 liters
1.65 gal / 6.25 liters
1.65 gal / 6.25 liters
Dimensions (in.)
24 x 14
3
/
8
x 26
1
/
2
24 x 14
3
/
8
x 26
1
/
2
24 x 14
3
/
8
x 26
1
/
2
h x w x d (cm)
61 x 36.5 x 67.3
61 x 36.5 x 67.3
61 x 36.5 x 67.3
Shipping Weight
155 lbs
165 lbs
170 lbs
Volts, Hz, Amps
120V, 60Hz, 6.8A
120V, 60Hz, 8A
120V, 60Hz, 13A
Volts Range
108V to 132V
108V to 132V
108V to 132V
60Hz Cat. No.
080501
080511
080521
Volts, Hz, Amps
240V, 50Hz, 3.4A
240V, 50Hz, 4A
240V, 50Hz, 6.5A
Volts Range
198V to 264V
198V to 264V
198V to 264V
Over Voltage: Category II
50Hz Cat. No.
080506
080516
080526
Additional Specifications for "P" Series 6000 Chillers with Positive Displacement Pump
Model "P" Series
6205P
6305P
6505P
Flow Rate @ 0 psi
1 gpm / 3.8 lpm
1 gpm / 3.8 lpm
1 gpm / 3.8 lpm
Pump Pressure (Adjustable)
20 to 100psi
20 to 100psi
20 to 100psi
Shipping Weight
160 lbs
170 lbs
170 lbs
Volts, Hz, Amps
120V, 60Hz, 9.6A
120V, 60Hz, 10.8A
120V, 60Hz, 15.8A
Volts Range
108V to 132V
108V to 132V
108V to 132V
60Hz Cat. No.
080502
080512
080522
Volts, Hz, Amps
240V, 50Hz, 4.8A
240V, 50Hz, 5.4A
240V, 50Hz, 8A
Volts Range
198V to 264V
198V to 264V
198V to 264V
Over Voltage: Category II
50Hz Cat. No.
080507
080517
080527
Additional Specifications for "T" Series 6000 Chillers with Turbine Pump
Model "T" Series
6205T
6305T
6505T
Flow Rate @ 0 PSI
3.5 gpm / 13.2 lpm
3.5 gpm / 13.2 lpm
3.5 gpm / 13.2 lpm
Pump Pressure (Adjustable)
20 to 100psi
20 to 100psi
20 to 100psi
Volts, Hz, Amps
120V, 60Hz, 9.6A
120V, 60Hz, 10.8A
120V, 60Hz, 15.8A
Volts Range
108V to 132V
108V to 132V
108V to 132V
60Hz Cat. No.
080503
080513
080523
Volts, Hz, Amps
240V, 50Hz, 4.8A
240V, 50Hz, 5.4A
240V, 50Hz, 8A
Volts Range
198V to 264V
198V to 264V
198V to 264V
Over Voltage: Category II
50Hz Cat. No.
080508
080518
080528
Environmental Conditions • Indoor Use Only
• Maximum Altitude : 2000 meter
• Operating Ambient : 5° to 30°C
• Relative Humidity : 80% for temperatures to 30°C
• Pollution Degree : 2
• Class 1 : Residential, Commercial, Light Industrial
• Class 2 : Heavy Industrial