11 function set to system error working mode, Advanced operations – Hach sc1000 User Manual

Page 110

Advertising
background image

108

Advanced Operations

6.3.3.11 Function set to SYSTEM ERROR working mode

SYSTEM ERROR

WARNING LIST

Default value: DISABLED
ENABLED: Monitors the internal warning bits of every probe.
DISABLED: Monitoring is disabled.

ERROR LIST

Default value: DISABLED
ENABLED: Monitors the internal error bits of every probe.
DISABLED: Monitoring is disabled.

SENSOR MISSING

Default value: DISABLED
Monitors the connection of every connected probe.
ENABLED: Monitoring is active.
DISABLED: Monitoring is not active.

ON DELAY
(0 s–999 s)

Default value: 5 seconds
Sets a delay for the relay to turn on.

OFF DELAY
(0 s–999 s)

Default value: 5 seconds
Sets a delay for the relay to turn off.

LOG INTERVAL

Default value: OFF
Sets the interval for logging the displayed value into the data logger.
Options: OFF, 5 min, 10 min, 15 min, 20 min, 30 min.