SECTION 2—OPERATING THE BATH LIFT

1900/1900S Bath Lift

12

Part No. 1148113

Emergency Lowering

NOTE: For this procedure, refer to FIGURE 2.6.

In the event of a handset or control box failure, the lift can still be lowered by using one 
of the three following methods, depending on reason the lift has not lowered:
If the battery is fine, but pendant is not or control box has failed and the lift needs 

to be lowered ‐ Press the tip of a ballpoint pen into the small opening below and to the 
right of the stop button on the control box (Detail “A”).

If the battery has failed and the lift needs to be lowered ‐ Pull down the red tag 

located above the battery near the push handle on the actuator column (Detail “B”). 

NOTE: Weight or force MUST be exerted upon the seat in order for this procedure to work.

FIGURE 2.6 Emergency Lowering

Disinfecting the Lift

NOTE: For this procedure, refer to FIGURE 2.7 on page 13.

ƽ WARNING

The seat and arms of the lift MUST be disinfected after each use with a bath.
ALWAYS wear protective gloves and eye protection when working concentrated

chemical disinfectants.

1. Prepare a working solution of Invacare disinfectant by doing the following:

A. Mix concentrated disinfectant with water to a dilution ration of half‐an‐ounce of 

disinfectant per one gallon of water.

B. Transfer the diluted disinfectant into a labeled spray bottle.

2. Position the bath lift into the tub. Refer to STEPS 13 to 16 in Transferring into/out of 

the Bathing Unit on page 9.

3. Spray both sides of the seat surface, the safety strap, the shoulder strap, and the arm 

rests of the lift. Refer to Detail “A” of FIGURE 2.7.

4. Gently wipe the push handles, scale, hand pendant, and any other surfaces that were 

handled during the transport and bathing of the resident with a wash cloth containing 
the diluted disinfectant. Refer to Detail “B” of FIGURE 2.7.

5. Let the disinfectant sit for ten minutes.

Press Ballpoint Pen Tip Here

for Emergency Lowering

DETAIL “A”

Control Box

DETAIL “B”

Pull Down Red Tag

for Emergency

Lowering

Actuator

Column