SM8000 Series Programming

Trigger Advance Register – Write Only

ADDR

Plug-In LA+0x3E

D15-D0 Unused

The act of writing to this location causes a Trace Advance event to
occur in the module. The specific data written to these bits has no
affect.


Board Busy Register – Read Only

ADDR

Plug-In LA+0x3E

D15-D7 Unused These

bits

always read back as 1s.

D6

Indicates whether the SMIP platform is a single or double-slot.
0 = single-slot
1 = double-slot

D5

A 0 read from this bit indicates the relays on module 5 have
settled. A 1 indicates that the relays on module 5 are still changing
state.

D4

A 0 read from this bit indicates the relays on module 4 have settle.
A 1 indicates that the relays on module 4 are still changing state.

D3

A 0 read from this bit indicates the relays on module 3 have
settled. A 1 indicates that the relays on module 3 are still changing
state.

D2

A 0 read from this bit indicates the relays on module 2 have
settled. A 1 indicates that the relays on module 2 are still changing
state.

D1

A 0 read from this bit indicates the relays on module 1 have
settled. A 1 indicates that the relays on module 1 are still changing
state.

D0

A 0 read from this bit indicates the relays on module 0 have
settled. A 1 indicates that the relays on module 0 are still changing
state.


Reserved Registers – Read and Write

ADDR

N/A

D15-D0 Unused

Writing to these registers has no effect and will always read back
as FFFF