Installation and Configuration

Basic Setup Procedure

Release 2.1

31

Basic Setup Procedure

This procedure can be used as a quick reference for those familiar with the
configuration requirements for the GPIB-M interface as installed in the DC power
supply.

IEEE-488 Controller Connection
Connect the GPIB cable to the supply at connector on the rear panel. Use an
approved IEEE-488 connector and cable when connecting the GPIB Interface to
your IEEE-488 GPIB network. Refer to

Figure 2.4

,

Figure 2.5

,

Figure 2.6

or

Figure 2.7

.

Add a ferrite block to reduce radiated emissions. The one inch square ferrite block

with built-in housing clip is packaged and shipped with the power supply interface

card.
To install the ferrite block:
1. Position the block no more than 5 cm (2 in.) from the power supply end of the

user cable.

2. Open the ferrite block housing.
3. Run the cable through the ferrite block. See

Figure 2.9, “XFR GPIB Cable with

Ferrite Block” on page 31

.

4. Close the housing clip.
The ferrite block ensures that the power supply system meets radiated emission

requirements for CE mark.

Figure 2.9 XFR GPIB Cable with Ferrite Block

GPIB Cable

Ferrite Block

To personal computer GPIB

connector

To power supply

GPIB connector