Installation and Configuration

Configure for Multichannel Operation

Release 2.1

39

Broadcasting

Commands

You can specify a channel address of "0" to broadcast a command to all power
supplies connect via CANbus. Broadcast commands are useful in set up and
operation.

Example:

Set all power supplies to remote mode:

SYST0:REM:STAT REM

Turn output on:

OUTP0 ON

Set voltage output to maximum:

SOUR0:VOLT MAX

Attempting to broadcast any query will generate an error.

Note that there will be a lag in execution time between the local unit and all other
units of up to a maximum of 20 ms.

Specifications

Note

Broadcasting may be useful in configuring a CAN-only power supply when its

address is unknown. Connect the unit to a GPIB-M power supply. Broadcast the set
multichannel command to change the CAN-only supply's address. The command has
no effect on the GPIB controlled unit.
Example:

"SYST0:COMM:MCH:ADDR 2"

will set the CAN-only unit to address 2.

Max connected units

50

Max cable length

40 m

Bus speed

700 kbits/sec

Termination

120 ohm 1/4 W

Connectors

modular handset 4-position, 4-contact jacks

Addresses

1 to 50