AVSL 600.100 MTM01 User Manual

Page 5

Advertising
background image

MTM01 User Manual


Multitester comparison table

Model

DCV

ACV

DCA

ACA

Ω

hFE

Cap

MTP01

Y

Y

Y

Y

Y Y

Y

Y

Y

Technical Specifications

Accuracies are guaranteed for 1 year, 23ºC ± 5ºC, less than 80% RH.

DC Voltage

RANGE

RESOLUTION

ACCURACY

200mV

100uV

± (0.5% of rdg + 3D)

2V

1mV

20V

10mV

± (0.8% of rdg + 5D)

200V

100mV

600V

1V

± (1.0% of rdg + 5D)


Input Impedance: 10MΩ

Overload Protection: 1000V DC or 750V AC rms

Max. Input voltage: 1000V DC

AC Voltage

RANGE

RESOLUTION

ACCURACY

200V

100mV

± (2.0% of rdg + 10D)

500V

1V

± (2.0% of rdg + 10D)


Input Impedance: 10MΩ

Frequency Range: 40Hz ~ 400Hz
Overload Protection: 1000V DC or 750V AC rms

Response: Average, calibrated in rms of sine wave
Max. Input voltage: 750V AC rms

Advertising