OWNER’S MANUAL

APPENDIX B

Dimensions and Clearances and Installation Requirements

B.1 The

AN74

i/ix Sterilizer consists of three major components: (1) a lower cabinet which

is used to contain the items being sterilized; (2) an upper cabinet containing the controller, user
interface display and pneumatic equipment to ventilate the lower cabinet and sterilization bag;
and (3) a 3/4” (1.9 cm) I.D. exhaust tube and fittings which vent the Ethylene Oxide laden air
from the lower cabinet and sterilization liner bag to the outside.

B.2

The total weight of the AN74 i/ix is approximately 50.5 pounds (22.9 kg) / 60.5 pounds

(27.5 kg). It is advisable to have two persons assist in the installation or moving of the ma-
chine. Refer to installation instructions in Appendix C and following for detailed instructions.
Place the sterilizer cabinet on a sturdy counter space or workbench at a height convenient for
loading.

B.3 CAUTION!: The Anprolene sterilizer cabinet location must be adjacent to an U

N

-

SWITCHED

,

GROUNDED

power outlet, preferably one that is dedicated to its exclusive

use. The standard Anprolene sterilizer is fitted with a 7 foot long (220 cm) power cord.

B.4. Whenever possible, the Anprolene sterilizer should be located against an outside
wall, and exhaust directly to the outside by means of 3/4” (1.9 cm) I.D. tubing to minimize
the impedance of the exhaust system.

CAUTION!: The exhaust outlet should not be located near a common area.