Flowchart for Ventilation Pump Failure During Sterilization Cycle

12 (or 24) HOUR CYCLE

STERILIZING

1:24:15 REMAINING

STERILIZATION HALTED

VENT PUMP FAILURE

1:24:15 REMAINING

ALARM OFF

STERILIZATION HALTED

VENT PUMP FAILURE

1:21:55 REMAINING

UNLOAD STERILIZER

STERILIZATION HALTED

VENT PUMP FAILURE

EXIT

If a VENT PUMP FAILURE occurs during
the sterilization cycle, the screen will display
the error message, and the machine will beep
continuously until the ALARM button is
pressed. This alerts the operator to the prob-
lem. The purge pump will start and remain on
for the remainder of the normal cycle time.

VENT PUMP FAILURE

REFER TO ERROR

SECTION OF MANUAL

In this case, sterilization has been stopped dur-
ing the cycle and the Ethylene oxide in the
liner bag was removed by the purge pump.

ITEMS IN THE LINER BAG SHOULD NOT
BE CONSIDERED STERILE

Refer to the Error and Troubleshooting sections
of the manual for information about repairing the
sterilizer

To reset the machine, shut off the machine power.
Then, while holding both buttons, turn on the
power. Release buttons after the display shows
characters.

The machine may be unloaded once the
counter reaches 0:00:00, but the items in the
liner bag should be resterilized prior to use

AN74 I

ANPROLENE STERILIZER