T-RAD-DB15--USB Instruction Manual

Version 2.1

December. 2012

16

VER

Queries the Firmware Version. No arguments are used.

Example.

Send: verCRLF

Reply: 1.00.00CRLF

RNG

Sets or Queries the instrument range

Example.

Send rng27CRLF sets the range to index 27, the 30mW range. See table
below for ranges versus indices

rngCRLF queries the current range.

Example

rngCRLF returns the range index

Reply 27CRLF

STR

Starts or stops the data flow.

Example.

Send str1CRLF The T-Rad will send data to the com port. Data is sent at
5Hz for power.

Send str0CRLF T-Rad will stop sending data and reply Ok.

Data is the hexadecimal data string. See the note below for how to decode
this value.

MAX

Query only. Returns the maximum range index for the head in use.

Example

Send maxCRLF

Reply 27CRLF

MIN

Query only. Returns the minimum range index for the head in use.

Example

Send minCRLF

Reply 20CRLF

ZRO

Command only. Sets the DC offset of the signal to the desired value.