T-RAD-DB15--USB Instruction Manual

Version 2.1

December. 2012

17

Example

Send zroCRLF

Reply 27,2350,2048CRLF

The values used are for example only. 2048 counts represent 1.25V of offset.
All heads use this value. 0 counts is 0V, 4095 counts is 2.5V. The reply is
Offset DAC0, Offset DAC1 code, ADC counts.

MRD

Query only. Returns the maximum reading for the head in use.

Example

Send mrdCRLF

Reply 216.7E-6CRLF, The maximum reading is 216.7uW.

TAU

Sets or Queries the time constant used by the Lock In Amplifier. The
argument is in seconds.

Example.

Send tau1.3CRLF sets the time constant to 1.3 seconds

tauCRLF queries the current time constant.

Example

tauCRLF returns the current time constant.

Reply 1.3CRLF

USN

Sets or Queries the User Serial Number

Example.

Send usnMySerialCRLF sets the user serial number to MySerial

This value is stored in Flash and allows a unique user specified serial number
to be assigned to each instrument.

usnCRLF queries the user serial number

UCD

Sets or Queries the User Calibration Date

Example.

Send ucdMM/DD/YYYYCRLF sets the local calibration date to
MM/DD/YYYY. The date must be in this format.

This value is stored in Flash and allows a unique user specified calibration
date to be assigned to each instrument.

ucdCRLF queries the user calibration date