Removing the display module – ZOLL X Series Monitor Defibrillator Rev D User Manual

Page 78

Advertising
background image

C

HAPTER

3

D

ISASSEMBLY

P

ROCEDURES

72

9650-001356-01 Rev. D

Removing the Display Module

Tools Required

#1 Phillips screwdriver with torque capability

3M copper adhesive tape

6mm socket wrench

Mid-Frame LCD Cable Routing Assembly Fixture, 9100-3028-AF (Optional)

Setup

This procedure should be conducted on a static mat.

Wear a wrist strap to ground during the procedure.

Complete these procedures in the following order before performing this procedure:

1.

“Removing the Handle” on page 63

2.

“Removing the Printer Module” on page 64

3.

“Disassembling the Main Unit” on page 66

4.

“Removing the Front Enclosure Module” on page 70

To reinstall the display module, reverse the steps.

Display Disassembly

1

Remove the grounding screw
and washer for EtCO

2

in the

area illustrated.

Advertising