Removing the cp carrier module – ZOLL X Series Monitor Defibrillator Rev D User Manual

Page 91

Advertising
background image

Removing the CP Carrier Module

9650-001356-01 Rev. D

X Series Service Manual

85

Removing the CP Carrier Module

Tools Required

#1 Phillips screwdriver with torque capability

Setup

This procedure should be conducted on a static mat.

Wear a wrist strap to ground during the procedure.

Complete these procedures in the following order before performing this procedure:

1.

“Removing the Handle” on page 63

2.

“Removing the Printer Module” on page 64

3.

“Disassembling the Main Unit” on page 66

4.

“Removing the Display Module” on page 72

5.

“Removing the EtCO2 and NIBP Modules” on page 76

6.

“Removing the Monitor Board and SpO2 Modules” on page 80

CP Module Disassembly

1

Remove the shield by pulling
it upwards over the CP carrier
board.

Note: During reassembly, be
sure to route antenna wires
through slot (red circle) in the
shield. Secure the shield to
the board by aligning with
slots on the board.

Advertising