1309 Multipoint Sampler

LumaSense Technologies A/S

Page 33 of 42

RAM Flag

Indicates that the RAM (random access memory) of the 1309, which

stores the 1309’s set-up data, has been corrupted. The flag is reset by
switching the 1309 off and then on. If the error persists, have the 1309

serviced.

PROM Flag

Indicates that the CRC (cyclic redundancy check) for the PROM (pro-

grammable read-only memory) has identified incorrect data. The flag is
reset by switching the 1309 off and then on. If the error persists, have

the 1309 serviced.

Job Specification Flag

Is set if an interface job sent to the 1309 is not recognised. This can be

due to an incorrect job header, incorrect syntax, or incorrect or missing
data. A job cannot be carried out by the 1309 if this flag is set. The
job’s syntax should be checked carefully and the job sent again. The

flag is reset by reading-out the Error Flags, or by switching the 1309 off
and then on.

Software Error Flag

Is set when the 1309’s software develops an error when running. When
this flag is set, the 1309 is reset automatically. The flag is reset by

reading-out the Error Flags, or by switching the 1309 off and then on.
If the error persists, have the 1309 serviced.

Power up Flag

This flag is set when the 1309 has been switched off and then on. This
flag is reset by reading-out the Error Flags. If the error persists, have

the 1309 serviced.