Chapter 5

______________________________________________________________________

_____________________________________________________________________________

BE6028-14

3434i SF6 Leak Detector

LumaSense Technologies A/S

Page 69 of 191

current through the contacts must not exceed 100mA. AC voltages

must not be connected to the Alarm Relay socket.

Fig.5.2 Configuration of the pins in the alarm relay socket

5.6 Continuous Warm System

The Continuous Warm System ensures that the monitor is kept warm

at any time that is also when the Monitor is not measuring.

If the Monitor is powered up at all time you can save the warm up

time when starting a measurement.

5.6.1

Setting-up the Monitor to Continuous Warm System

The procedure to enable the Continuous Warm System is as follows:

1. Press the following push-button sequence: SET-UP S3 S1 S1,

and the following text is displayed:

2. Press S3 q three times S3 SET-UP. This selects the

continuous warm system.

3. Press RESET S1 (Partial Reset) in order to enable the continuous

warm system.

SELECT GENERAL SET-UP BRANCH ?

CLOCK SOUND TESTS

Alarm relay 1

Alarm Relay 2