Metex 2200P pH Analyzer/Controller User Manual

Page 23

Advertising
background image

Model 2200P

Page 21

5.13 Temperature

5.14.1 The temperature of the process can be read at any time by entering the menu and

scrolling through the menu until TEMPERATURE LED is illuminated. Either °C or
°F will be indicated depending on the position of DIP switch No.1 of Bank S1. If the
temperature is suspected of being incorrect, place the S41 switch on the back of the
main circuit board, in the up position, (this places a 300 ohm across the BL and YL
terminals) the temperature should read 25°C or 77°F. If not return S41 to down
position and re-calibrate the temperature as per Section 7.4.

5.14.2 The 0-5 Vdc and 0-1 mA analog outputs can be selected to track the process

temperature by simply placing DIP switch No. 8 of Bank S1. in the OFF position.
The temperature span of the output is set to the utility menu. See Section 7.4.

5.14 Status

5.15.1 The 2200P continuously checks the integrity of all stored data and monitors the

condition of the measuring system. If a fault is detected, the STATUS LED above
the display will change from green to red. When this occurs, enter the menu and
scroll through the menu until the STATUS LED illuminates. The display will show a
numerical code to assist you, in correcting the problem.

The following table shows the display codes, causes and remedies:

CODE

POSSIBLE CAUSE

SUGGESTED REMEDY

0

Normal Condition

No action required

1

Reserved

2

Temperature Sensor off Scale

Check for open or short
connections of temperature sensor
input (Terminals BK & YL on TB2)

3

PH reading off Scale

Verify process. Check for open or
short connections

4*

Memory Loss

Call your AquaMetrix Rep. Or
AquaMetrix directly

5

Reserved

6

Calibration out of limit

Check scale setting (See range
change in Utility Menu)

7

Factory settings in force as a result
of the “Factory Reset Procedure”

8

PH calibration not correct, or not
performed

Re-calibrate (See Sec. 5.2 & 5.3)

NOTE: Code 4 could be serious failures so the alarm relay will activate, in addition to the status

LED changing to red, if programmed by the position of DIP switch No. 5 of Bank S1. See
Section 4.7.

Advertising