0 specifications – Metex 2200P pH Analyzer/Controller User Manual

Page 4

Advertising
background image

Model 2200P

Page 2

2.0

SPECIFICATIONS

DISPLAY:

4 x 7 segment _” LED Display

MEASURING RANGES:

pH: 0.01 to 14.00 pH
Temperature: 0°C to 100°C (32°F to 212°F)

POWER REQUIREMENTS:

Standard: 98-132 Vac, 50/60 Hz (less than 20 VA)
Optional: 196-264 Vac, 50/60 Hz (less than 20 VA)

23-26Vdc (nominal current 150mA)

AMBIENT CONDITIONS:

-30 to 50°C (-22 to 122°F)
0 to 90% R.H. non-condensing

CONTROL RELAYS:

5A 115/230 Vac, 5A 30 Vdc SPDT
Selectable to activate in response to rising or falling
process
Fully adjustable relay on and relay off setpoints
Manual Override Auto / Off / On for Relay A
Cycling Feature: On time 0-600sec off time 0-600sec
Normal or Fail Safe operation

ALARM RELAY:

5A 115/230 Vac, 5A 30 Vdc SPDT
Programmable to activate at high and low alarm
setpoints
Deadband fixed at 2% of Full Scale
Normal or Fail Safe Operation

RELAY INDICATORS:

Three individual LEDs indicate status of the two
control relays and the alarm relay.

ANALOG OUTPUTS:

Non Isolated 0-1mA, 100 ohms maximum load
Non Isolated 0-5Vdc, 1000 ohms minimum load
Isolated 4-20mA, 800 ohms maximum load
Range Expand: The analog outputs can be scaled to
represent any segment of the full scale of the
instrument. Minimum segment is 10% of Full Scale.
Output Hold: The analog outputs are automatically
placed on hold when the unit is placed in any menu
mode.
Temperature Output: The 0-1mA / 0-5 Vdc outputs can
be selected to track the process temperature or pH.

SENSOR-TO-ANALYZER DISTANCE:

Differential Probe (P60) 3000 feet maximum
Conventional combination probe 25 feet (an
AquaMetrix Series 101 preamplifier is required for
distances greater than 25 feet)

CALIBRATION:

Two point or automatic with buffer recognition.

TEMPERATURE COMPENSATION:

Automatic 0°C to 100°C (32°F to 212°F) If probe has
temperature compensation

SYSTEM ERROR:

A LED indicates a system alarm. The alarm relay can
be programmed to activate upon system memory loss.

DIAGNOSTICS:

Invalid entries are identified by respective flashing
LEDs. When a status error occurs, press CALL to
obtain the status error number, which corresponds to a
fault as described in the Status Fault Table.

TEST:

Display value and analog outputs can be set manually
to any value for testing and diagnostic purposes. This
feature allows the operation of the control relays,
alarm relay, and outputs to be tested independently of
process.

SAFETY AND SECURITY:

Non-volatile memory (EEPROM)
Password protected if enabled
Watch-dog timer monitors microprocessor
Instrument automatically returns to on line operation if
accidentally left in menu mode.

SENSITIVITY: 0.1% of span

STABILITY: 0.1% of span per 24 hrs. non-cumulative.

NON-LINEARITY: 0.1% of span

REPEATABILITY: 0.1% of span or better

TEMPERATURE DRIFT:

Zero: 0.01% of span per °C
Span: 0.01% of span per °C

RESPONSE TIME:

0.5 second to 90% of value upon step change

ENCLOSURE:

NEMA 4X molded fiberglass reinforced polyester
enclosure with four 1/2" conduit holes and mounting
feet for surface mount. A NEMA 4 plug is provided for
one hole.

MOUNTING CONFIGURATIONS:

Standard is surface mount
Optional panel mount hardware Part No. C35-68
Optional pipe mount hardware Part No. C35-69

NET WEIGHT: 3 1/2 lb. (1.6 kg)

Approvals: CSA