Chapter 6 ___________________________________________________Calibration and Adjustment

VAISALA ______________________________________________________________________ 111

CHAPTER 6

CALIBRATION AND ADJUSTMENT

This chapter contains instructions for performing the calibration and
adjustment of the product.

The PTB330 is fully calibrated and adjusted as shipped from the
factory. Calibration must be done always when there is reason to
believe that the device is not within the accuracy specifications.

When defining the calibration interval, you must the local long term
specifications and other requirements into consideration. Contact
Vaisala Service Centers for details.

It is recommended that calibration and adjustment should be carried
out by Vaisala. See sections Technical Support and Product Returns
on page 110. Calibration and adjustme
nt are carried out through the
serial port or with the optional display/keypad.

Pressure

You can select a simple offset or a two-point offset and gain
adjustment for the barometer module using the LCP command. The
MPCP

command is for the more sophisticated multipoint adjustment

with up to eight pressure levels.

NOTE

Entering new linear or multipoint adjustments cancel the previous
adjustments. It is advisable to write down the previous linear and
multipoint adjustments so that they are not lost by mistake.