Removing devices – Vaisala viewLinc 4.3 User Manual

Page 116

Advertising
background image

Removing Devices

108

4

|

4

M211679EN-A

To edit calibration properties:

1

Make sure a Device which supports calibration properties is
selected.

2

Ensure the initial calibration values have been downloaded
from the Device to viewLinc (automatically discovered).

3

On the viewLinc desktop choose, Options | System Con-
figuration | Locations Manager

, and then use the Devices tree

to select a Channel to edit.

4

On the Options | Channel menu select

 Edit Channel

(or use

the right-click menu).

5

In the Edit Channel Properties window, edit the calibration
properties for Calibration Scale and Calibration Offset
(provided by Vaisala or collected from your onsite calibration
testing).

6

Click OK.

7

When the calibration property is changed, confirm the change
and ensure the correct calibration date is entered.

8

Click OK.

Removing Devices

When a Device is removed from the viewLinc system, or transferred
to a new monitoring area, viewLinc needs to be notified to prevent
system alarms.

You may want to remove a Device when:

a Device requires maintenance (such as recalibration)

a Device is no longer required

a Device needs to be disabled temporarily

To ensure consistency in monitoring, you can choose to deactivate a
Device, deactivate a Location, swap a device or permanently delete
a
Location. Before you can swap a Device you must first unlink any
linked Channels.

IMPORTANT: If a Device is removed without being deactivated,
this will trigger a Communication Alarm.