When the sensor has been installed, the user may need to change some
of the default settings. For details, see section Initial Settings on page
56.

Table 13 below lists the settings and the corresponding commands.

Table 13

Settings and Corresponding Commands

Operation

Command

Baud rate

BAUD

Polled or automatic mode, message type
setting

AMES

Sensor ID

CONF

Weather algorithm parameters

WSET

Table 14 below lists the routine maintenance commands.

Table 14

Routine Command for Maintenance

Operation

Command

Sensor cleaning

CLEAN
(optional)

Visibility calibration

CHEC, CAL

Temperature calibration

FREQ, TCAL

Weather algorithm parameters

WSET

The standard output messages contain a status character, which
presents the results of the internal diagnostics to the host computer or
the user. If the sensor indicates a warning or an alarm in a standard
output message, the host computer or the user can obtain a detailed
status report by using a special command (STA). The status report can
also be polled (message 3) in place of the standard data message.
Usually, the detailed status information is sufficient for locating the
fault.

Table 15 below lists the status report command.

Table 15

Status Report Command

Operation

Command

Getting a status report

STA