background image

Chapter 8 ____________________________________________ Sensor and Data Message Settings

VAISALA ______________________________________________________________________ 103

Precipitation Sensor

Checking the Settings

With this command you can check the current precipitation sensor
settings.

Command format in ASCII and NMEA 0183: aRU<cr><lf>

Command format in SDI-12: aXRU!

The response in ASCII and NMEA 0183:

aRU,R=[R],I=[I],U=[U],S=[S],M=[M],Z=[Z]<cr><lf>

The response in SDI-12:

aXRU,R=[R],I=[I],U=[U],S=[S],M=[M],Z=[Z]<cr><lf>

where [R][I][U][S][M][Z] are the setting fields, see the following
section.

Example (ASCII and NMEA 0183, device address 0):

0RU<cr><lf>

0RU,R=11111100&10000000,I=60,U=M,S=M,M=R,Z=M<cr><lf>

Example (SDI-12, device address 0):

0RU!0RU,R=11111100&10000000,I=60,U=M,S=M,M=R,
Z=M<cr><lf>

0RU!0RU,R=11111100&10000000,I=60,U=M,S=M,M=R,Z=M<cr><lf>

where
a

=

Device address

RU

=

Precipitation sensor settings command in ASCII and
NMEA 0183

XRU

=

Precipitation sensor settings command in SDI-12

<cr><lf>

=

Command terminator in ASCII and NMEA 0183

!

=

Command terminator in SDI-12