Aalborg Vortex User Manual
Page 92
87
COMMAND
NAME
DESCRIPTION
NO.
COMMAND SYNT
A
X
COMMAND
ARGUMENT 1
ARGUMENT 2
ARGUMENT 3
ARGUMENT 4
RESPONSE
Pulse
Output
Sets and controls action of the
programmable Pulse Output
cir
cuitr
y.
NOTE: Unit/Pulse
value has to be entered:-for
V
o
lumetric flow conditions
in currently selected volume EU
-for Mass flow conditions
in currently selected mass EU
EU has to be not time based.
It is recommended to set the
unit/pulse value equal to the
maximum flow in the same
units per second. This will limit
the pulse to no faster than one
pulse ever
y second.
Example: Maximum flow rate:
1200 liter/min (1200 liter/min =
20 liters per second)
If Unit/Pulse is set to 20 liters
per pulse, the output will pulse
once ever
y second.
13
P
C (flow
configuration)
V
-volum. flow
M-mass flow
<V
alue>
(Unit/Pulse)
V
o
lumetric flow -
current
V
o
lumetric E.U.
Mass flow - cur-
rent mass EU
(example:
10 litr/pulse)
PCV
:<value>,<value>
Example:
PCV
:V
,10
Or
PCM:<value>,<value>
Example:
PCM:M,2.5
T (Pulse active
T
ime in ms)
<value>
[10 -13000 ms]
PT
:<value>
Example: PT
:1000
D (disable pulse
output)*
P:D
E (enable pulse
output)
P:E
Q (read current
pulse output
Queue value)
PQ:<value>
(number of pulses)
S (setting status)
PS:Mode,FlowConditions,
Unit/Pulse,PulseT
imeInter
val
Fluid
Density for
standard
conditions
Read and set Fluid Density for
standard conditions in g/cm3
NOTE: it is not measured
density
.
14
FD
(New Density
V
alue)
0.0001 @
Density
@ 10.0 g/cm3
FD:<value>
Example:
FD:1.293
No Argument
(Returns Current
Density in g/cm3)
FD:<value>
Example:
FD:1.293