Aalborg Vortex User Manual
Page 97
92
COMMAND
NAME
DESCRIPTION
NO.
COMMAND SYNT
A
X
COMMAND
ARGUMENT 1
ARGUMENT 2
ARGUMENT 3
ARGUMENT 4
RESPONSE
Calibration
Settings
Sets/Reads
Calibration
related variables.
Argument 1 = T
Hours since last
time unit was
calibrated.
NOTE: has to be
reset to zero after
calibration.
19
CS
S
Meter Size/Pipe ID
[inches]
<V
alue>
[inches]
CSS:<value>
Example: CSS:0.75
No Argument (Returns
Current Meter Size settings)
CSS:<value>
Example: CSS:0.75
F
Meter Full Scale
Range in liter/min
<V
alue>
[liter/min]
CSF:<value>
Example CSF:1000.0
No Argument (Returns
Current Meter FS Range)
CSF:<value>
Example: CSF:1000.0
C
Meter Calibration
Factor Hz/liter/min]
(for STD condi-
tions)
<V
alue>
[Hz/l/min]
CSC:<value>
Example CSC:1.0
No Argument
(Returns Current Meter
Calibration Factor)
CSC:<value>
Example CSC:1.0
O
Meter Low and
High Frequency Cut
Off [Hz]
<Low Cut off V
alue> (Hz)
<High Cut off
V
alue>(Hz)
CSO:<Low CutOff V
alue>,
:<High CutOff V
alue>
No Argument
(Returns Current Cut off
V
alues settings)
CSO:<Low CutOff V
alue>,
:<High CutOff V
alue>
L
Meter Low Flow
Cut Off in % of full
scale
<V
alue>
(%FS)
CSL:<value>
Example: CSL:5.0
No Argument
(Returns Current value)
CSL:<value>
Example: CSL:5.0
T
Calibration/Mainten
ance T
imer
No Argument
(read timer)
CST
:<value>
Example: CST
:1024.2
Z
Set Cal. T
imer to Zero
CST
:Z