33

Pressure

Source

PROCESS CALIBRATOR

QUIK-CHEK

READ PRESSURE

Choose this function to measure pressure with a ModPak 91
module holder and one of the many QuikCal 90 pressure modules.

1) Select the proper module for the pressure range of the process
2) Place the QuikCal 90 Module in the ModPak 91 holder
3) Plug the ModPak cable into the mating connector on the 830-KP
4) Turn the selector knob to PRESSURE
5) The 830-KP will display – – – – – for up to 20 seconds while

the module is powered up and verified

6) The 830-KP will display the range of the module for three seconds
7) Press the °C/°F/ZERO push-button to “Zero” the pressure

reading

Carefully connect the module to the pressure point to be
measured.

To Change the pressure engineering units
1) Press the TYPE/ENG UNITS push-button. The words SELECT

ENG. UNITS and the current selection appear on the display.

2) Press any UP/DOWN push-button until the required type

appears on the display.

3) Press the STORE/RESET push-button or wait 5 seconds to

store the selection

OPERATING INSTRUCTIONS

32

READ PRESSSURE

WARNING! - SAFETY CONSIDERATIONS
Sudden release of compressed or stored gas can cause personal
injury. Always vent the system before making pressure
connections or disconnections. Exercise standard physical
protection practices such as eye protection, gloves, protective
clothing, etc.

To prevent the potentially hazardous release into the atmosphere
of substances introduced into the pneumatic system by the user,
no relief valves are provided in the QuikCal 90 Pressure Module.
Consequently, if a module is overpressurized, the sensor will be
damaged.

FITTING INSTALLATION
All pneumatic connections to the QuikCal 90 Module are made via
1/8”-27 NPT threads in the pressure port. Adapter fittings may be
used for 1/4”-18, straight 10-32 threads or any other size fittings to
match the field connections.

To install a fitting in a module pressure port:
1) Apply Teflon tape or pipe dope to the fitting threads (Teflon tape

not recommended with stainless steel fittings)

2) Carefully install the fitting in the pressure port on top of the

module

3) Use a 5/8” wrench to support the pressure module port and a

second wrench to tighten the fitting

4) Check for leaks

OPERATING INSTRUCTIONS