7.12.3 Perform A Single Level Calibration

To quantify unknown samples, the method must contain the peak areas for each
component in the standard calibration sample. To acquire the peak area
information:

Run the standard calibration sample and designate it as a calibration run
following the steps below, or

Calibrate the method using a stored run

To calibrate:

1

Click Analysis >> Analysis/Single Level Calibration. See

Figure 7-47

.

Figure 7-47 Menu bar location for analysis/single level calibration