2.2 Environmental Conditions

The following specifications are applicable to the 3000 Micro GC Heated Regulator
and Heated Vaporizer sample conditioners.

Operating temperature range. . . . . . 0 to 50°C

Relative humidity . . . . . . . . . . . . . . . 5 to 95% (non-condensing)

Altitude . . . . . . . . . . . . . . . . . . . . . . . 4572 m (15000 ft.)

Usage location . . . . . . . . . . . . . . . . . Temperature controlled environment

2.3 Instrument Specifications

2.3.1 Sample Inlet Specifications

Sample type . . . . . . . . . . . . . . . . . . . Gas

Maximum sample pressure . . . . . . . <172 kPa (25 PSI)

Recommended sample pressure . . . Ambient to 69 kPa (ambient to 10 PSI)

2.3.2 Sample Injector Specifications

Injector types

Variable volume/timed, heated

Variable volume/large loop, heated

Fixed volume, heated

Backflush to vent, heated

Injection volume

1 to 10 μL for variable volume/timed

1 to 30 μL for variable volume/large loop

1 μL for fixed volume injector and backflush injector with fixed mode

2.3.3 Column Heater Specification

Isothermal Operation . . . . . . . . . . . . Ambient plus 15 to 180°C

2.3.4 TCD Specifications

Internal Volume . . . . . . . . . . . . . . . . 240 nL

Detection Limit . . . . . . . . . . . . . . . . . Low ppm to 100%

Linear Dynamic Range. . . . . . . . . . . 10

6

±10%