Troubleshooting guide – Labnet Z100A Clinical Lab Centrifuge User Manual

Page 8

Advertising
background image

7

6.

Troubleshooting guide

Centrifuge will not start

Possible reason:

No power supply.

Solution:

Check main power supply and centrifuge fuse.

Possible reason:

Blown fuse.

Solution:

Check fuse and replace if necessary (see 5.5)

Lid cannot be released:

Possible reason:

Lid lock is jammed.

Solution:

Push front of lid down while pushing release button.

Centrifuge cannot be started, although power is on:

Possible reason:

Lid not closed correctly.

Solution:

Close lid correctly, push down on front of lid.

Possible reason:

No speed or time has been selected.

Solution:

Check and set the time or speed.

Possible reason:

Blown fuse.

Solution:

Check fuse and replace if necessary (see 5.5).

Should you have any questions about the centrifuge and its accessories, or require
service, please contact the Labnet International Service Department at 1-888 522-
6381. Please have the unit’s serial number (located on the back panel of the instru-
ment) available when calling. Do not send in a unit for service without first calling to
obtain a repair authorization number and a decontamination form. Should the cen-
trifuge need service be sure to decontaminate it and enclose the decontamination
from. The unit should be properly packed to avoid damage. Any damage resulting
from improper packing shall be the responcibility of the user.

Advertising