Figure 4: Model 340 front panel controls and indicators

MODE SELECTOR switch and indicators
The three position switch selects the calibration mode. LEDs indicate the current
mode.

• ZERO: The permeation chamber output is dumped out the chamber vent;

only the dilution stream exits through the stream outlet to the analyzer.

• SPAN: The permeation chamber output and the dilution stream are mixed

and fed through the stream outlet to the analyzer.

• REMOTE: Unit defaults to the ZERO mode. A remote contact closure to rear

panel terminals C and V1 will switch the unit to the SPAN mode.