Configuration 5

in this configuration (Figure 10), pump P1 draws gas in through the supply
inlet on the back panel and delivers it to accumulator AC1. The accumulator
smoothes out flow fluctuations generated by the pumping action and provides
a port for releasing pump backpressure through pressure relief valve RV1, set at
5 psig. The tee at the output of the accumulator splits the supply into carrier and
dilution streams, each of which is plumbed to a back-panel outlet port which
must be plugged if an external supply is used. These ports can be connected
individually through external scrubbing media (filters, driers, etc.) or simply
looped to their respective inlet ports. Filters F1 and F2 remove particulates
larger than 5 microns from these two streams.

SUPPLY

INLET

DILUTION

STREAM

CARRIER

STREAM

TEE

MAIN PUMP

P1

ACCUMULATOR

AC1

BACK PRESSURE

REGULATOR

BPR1

FILTER

F1

DILUTION

OUTLET

DILUTION

INLET

BYPASS

LOOP

FILTER

F2

CARRIER

OUTLET

CARRIER

INLET

BYPASS

LOOP

Advanced Theory of Operation

Figure 10: Diagram of Configuration 5