VICI 450 Dynacalibrator User Manual

Page 6

Advertising
background image

2

Introduction

As indicated in Figure 1, the carrier stream passes through scrubber #1 and
micron line filter LF1 to the differential pres sure regulator DPR1. The scrubber
contains spe cially-activated charcoal designed for broad-based scrubbing action.
This scrubbing medium is particularly effective in removing sulfur dioxide and
other sulfur compounds.

The differential pressure regulator and carrier flow restrictor FR1 together
provide a highly stable carrier stream flow rate to the permeation chamber.
The standard carrier flow restrictor consists of a disk with a 0.006-inch diameter
orifice. The flow rate through the orifice is a function of the pressure differential
across it and this differential is maintained at 3.2 psi by DPR1.

SAMPLE

INLET

DILUTION

STREAM

INPUT

TEE

TEE

MIXING

TEE

OVERFLOW

VENT

CARRIER

STREAM

INPUT

SCRUBBER

#1

SCRUBBER

#2

LINE FILTER

LF1

DIFFERENTIAL

PRESSURE

REGULATOR

DPR1

CARRIER

FLOW

RESTRICTOR

FR1

PERMEATION

CHAMBER

CHAMBER

FLOWMETER

FM1

SPAN 1

FLOWMETER

FM2

TEE

DIFFERENTIAL

PRESSURE

REGULATOR

DPR2

SPAN 1 FLOW

CONTROL VALVE

FCV2

SOLENOID

CONTROL

VALVE

SV2

CHAMBER

VENT

SPAN 2

FLOWMETER

FM3

NO

NC

C

NO

NC

C

SPAN 2 FLOW

CONTROL VALVE

FCV3

SOLENOID

CONTROL

VALVE

SV1

TEE

NO

NC

C

SOLENOID

CONTROL

VALVE

SV3

STREAM

OUTLET

Figure 1: Model 450 plumbing schematic

The output of the permeation chamber is fed to solenoid control valve SV2.
When the front panel mode selector switch is in either the SPAN 1 or SPAN 2
position, the chamber output is routed through the valve from its C to its NC port
and to the mixing tee, where the calibrated dilution stream is added and mixed
into it to set the final concentration. When the front panel mode selector switch
is at STANDBY, ZERO, or REMOTE, SV2 is de-energized and the chamber output
passes through its C and its NO ports to the chamber vent. The dilution stream
alone then passes through the mixing tee.

The output of the mixing tee is split by another tee and fed to the overflow vent
and solenoid control valve SV3. When the front panel mode selector switch is in
either the STANDBY or REMOTE position with no remote inputs, SV3 blocks the
stream and causes it to be dumped to the rear panel overflow vent. The analyz-
er’s input is routed through the rear panel sample input port, the NO-to-C ports
of SV3, and the rear panel stream outlet.