9.4 TROUBLESHOOTING

A reference guide to common operating problems with the centrifuge is presented below.

23

PROBLEM

CAUTION

ACTION

Centrifuge shows no sign

of power, but primary

electrical supply is available.

Loose power connection.

Check all power plugs.

Centrifuge does not operate

when the ON button is pressed.

Lid is not latched.

Close and latch lid.

Failure of micro-switch

in latch mechanism.

Request service.

Centrifuge lid will not open,

but primary electrical supply

is available.

Centrifuge rotor is spinning.

An electrical interlock prevents

the lid from opening until rotor

stops spinning. Wait until all motion

ceases.

Latch release delay.

Wait 4-5 s

after rotor stops.

Plunger of latch solenoid

is jammed.

See Section 9.3

and release solenoid plunger.

Centrifuge vibrates excessively.

(Note: momentary vibration is

normal during acceleration,

but should disappear upon reaching

full speed.)

Press power button OFF

to stop rotation.

See Action steps.

(Note: Centrifuge is capable

of 30º tilt without causing excessive

vibrations.)

a) Be sure tube load is properly

balanced.

b) Tighten rotor nut with wrench,

replace and tighten rotor cover.

Close top lid

and re-check for vibration.

c) Check that surface does

exceed 30º tilt limit.