9.5 TROUBLESHOOTING (continued)
The combinations of On, Off and Flashing status of the ‘POWER’ and ‘SPEED’ LED’s shown below provide visual
indicators of the current operating condition.
9.6 PARTS LIST
The following spare parts and accessories for the QBC Capillary Centrifuge are available from your QBC System
distributor or through the nearest QBC Diagnostics Inc. office:
Part Description
Catalog No.
Rotor Head
421289
Head Nut
421290
Rotor Cover
421291
Head Nut Wrench
421292
Rotor Removal Wire
4277-020-006
Line Cords, Power Pack:
United States
421634
Europe 421551
United Kingdom
421554
Power Pack
421763
Operator’s Manual
422928
Yellow LED
‘POWER’
Green LED
‘SPEED’
Centrifuge Status
Operator Action
Off
Off
Power off, no rotor motion.
Plug in Power Pack.
Flashing
Off
Ready to spin, lid latched.
Press ON button.
On
Off
Power on, no rotor motion,
lid not latched.
Load or unload tubes.
On
Flashing
Power on, lid latched,
rotor under speed.
Rotor is usually accelerating
to design speed.
On
On
Power on, lid latched,
rotor at design speed.
No action.
Flashing
Flashing
Power on, lid latched,
overspeed shutdown.
Request service.