9.5 TROUBLESHOOTING (continued)

The combinations of On, Off and Flashing status of the ‘POWER’ and ‘SPEED’ LED’s shown below provide visual
indicators of the current operating condition.

9.6 PARTS LIST

The following spare parts and accessories for the QBC Capillary Centrifuge are available from your QBC System
distributor or through the nearest QBC Diagnostics Inc. office:

Part Description

Catalog No.

Rotor Head

421289

Head Nut

421290

Rotor Cover

421291

Head Nut Wrench

421292

Rotor Removal Wire

4277-020-006

Line Cords, Power Pack:

United States

421634

Europe 421551
United Kingdom

421554

Power Pack

421763

Operator’s Manual

422928

Yellow LED

‘POWER’

Green LED

‘SPEED’

Centrifuge Status

Operator Action

Off

Off

Power off, no rotor motion.

Plug in Power Pack.

Flashing

Off

Ready to spin, lid latched.

Press ON button.

On

Off

Power on, no rotor motion,

lid not latched.

Load or unload tubes.

On

Flashing

Power on, lid latched,

rotor under speed.

Rotor is usually accelerating

to design speed.

On

On

Power on, lid latched,

rotor at design speed.

No action.

Flashing

Flashing

Power on, lid latched,

overspeed shutdown.

Request service.