Care and preventative maintenance – Drucker Diagnostics Model 842HS Centrifuge User Manual

Page 11

Advertising
background image

CARE AND PREVENTATIVE MAINTENANCE:

With proper care and maintenance your Horizon centrifuge will provide years of laboratory service.

For proper care, the following steps should be taken:

1.

Provide Adequate Ventilation: For cooling purposes, the Horizon draws in ambient air

through the air intake cover on the top of the lid and exhausts this air in the rear of the base.

The centrifuge should be placed on a hard smooth surface for good air circulation.

2.

Always Spin Balanced Loads: This centrifuge is equipped with imbalance detection.

However, it is still possible to run loads that are minimally imbalanced, particularly at an

imbalance setting of LOW SENSITIVITY. For maximum centrifuge life, set the imbalance

detection to HI SENSITIVITY and always run balanced loads. Refer to pages 8 and 9 on for

additional information on balancing the load.

3.

Keep the Rotor Clean: NOTE: Always follow the safety guidelines of your laboratory to

properly clean up and/or dispose of materials in the event that a substance known to be

potentially toxic, radioactive or contaminated with a pathogenic microorganism is spilt in

or on the centrifuge. Small glass fragments left in the rotor after a tube breakage

may adhere to the next test tube inserted in that position. When this tube is handled,

these fragments may puncture protective gloves and lacerate the operator’s fingers or hand.

Remaining fragments may provide stress points on subsequent tubes and result in additional

breakage. If a tube breakage occurs, carefully remove the tube. Properly dispose of the

sample and tube fragments and thoroughly clean the rotor.

4.

Motor and Electrical Maintenance: This centrifuge uses a brushless-DC motor. There

are no brushes to replace and it should not need routine servicing for the life of the

centrifuge. The electrical components are selected for high reliability and should not need

service.

5.

Keep the Centrifuge Clean: The cabinet, rotor top and accessories shall be thoroughly

cleaned using either isopropyl alcohol, soap and water. Must be Lysol (1/4 cup/gallon) or “Cidex”.

Not all Lysol will be sufficient (must be a Lysol Amphyl disinfective cleaner or Virkon). The use of

Fully/Partially Halogenated Hydrocarbons, Ketones, Esters and all other chemicals not prescribed

by the manufacturer may cause damage to the rotor and tube carriers / holders and shall not be

used. Before using any cleaning or decontamination methods except those recommended by the

manufacturer, users should check with the manufacturer that the proposed method will not

damage the equipment.

If it is necessary to remove the rotor for additional cleaning it is required that a qualified

technician remove the rotor assembly.

Apply cleaning solutions with a towel or cloth. Do not submerge the centrifuge in water or

other cleaning solutions as this will cause damage and void your warranty!

If hazardous material is spilt on or inside the centrifuge, the centrifuge must be decontaminated.

page 11

Advertising