Setup location, Initial setup procedure – Drucker Diagnostics Model 842HS Centrifuge User Manual

Page 6

Advertising
background image

SETUP LOCATION:

1. Unpack the centrifuge and verify that all of the supplied equipment is present.
2. Choose a setup location which meets the following criteria:
a) A bench top clearance height of 40 cm is required in order to open the lid.
b) The clearance envelope is the space around the centrifuge which is required for

safety. Choose a setup location which will allow for a clearance envelope of at least

61 x 61 cm, (with the centrifuge at the center). No person or hazardous material shall

be permitted in the clearance envelope during operation. The operator time within

the envelope shall be limited to the time necessary for loading, unloading and

centrifuge

operation

only.

c) Proper ventilation is necessary to prevent the overheating of samples as well as

premature failure of the centrifuge. Choose an area which will allow unencumbered

air flow.
d) The centrifuge is designed to rest on its four rubber feet. No adjustment is necessary

for leveling the centrifuge, however, the surface should be flat and level.
e)

Be sure the outlet is always within reach as the line cord is the means of

emergency

disconnection!

INITIAL SETUP PROCEDURE:

If any problems are found during the initial setup

procedure, refer to the troubleshooting section. For further assistance, contact The Drucker

Company. at 814-342-6205
1. Plug the female end of the line cord into the rear of the centrifuge. Plug the male end into

an approved electrical outlet. For electrical safety, the unit must always be properly

grounded.
2. For operator safety, the locking system is always active; requiring power and direction

from the user to disengage it (the lid also automatically unlocks at the end of a run when

it is safe to do so). To unlock the lid (in order to access the rotor chamber) press the

‘OPEN / STOP’ button on the control panel. The ‘UNLOCKED’ indicator light should

illuminate. If it does not, refer to the section on troubleshooting. The lid will remain

unlocked for 15 seconds after pressing the ‘OPEN / STOP’ button.
3. Turn the lock counterclockwise and open the lid.
4. Spin the rotor by hand; check for free and level rotation. If the rotor does not spin freely,

refer to the section on troubleshooting.
5. Turn rotor lid knob counterclockwise to remove.
6. Place the test tubes inside the rotor and verify that they are seated properly.

7. Place the rotor lid on the rotor and turn knob clockwise until snug.
8. Close the lid. Rotate the lid knob clockwise to its complete stop position. The

’LOCKED’ indicator light should be illuminated. If it is not, make sure that the lid is

locked properly. The centrifuge will not run unless the lid is latched properly and the

’LOCKED’ indicator light is illuminated.
9. Use the up arrow button next to the RPM display to set the speed to 12000 RPM.
10. Initiate a test run by pressing the ‘START’ button.
11. The ‘RUNNING’ indicator light will illuminate.
12. The unit will accelerate to the current set speed.
13. Listen to the sound of the centrifuge. A smooth whirring sound should be heard. If there

are any loud or unusual sounds, stop the centrifuge by pressing the ’OPEN / STOP’ button

immediately and refer to the section on troubleshooting.
14. Press the ‘OPEN / STOP’ button to terminate the test run. The rotor will decelerate to a

complete stop and the lid will then unlock automatically for sixty (60) seconds.
15. Take time now to familiarize yourself with the various additional user settings. Turn to

page 8.
16. The centrifuge is now ready for operation.

page 6

Advertising