PART NO. # MAX TUBE SIZE SPEED RANGE FORCE RANGE

6-Place Horizontal (Standard)

7786021

1

75mm tube holders

7713043

1

6 17mm x 75mm 1000 to 4,500 RPM 140 to 2,260 xg

100mm tube holders 7713040

1

6 17mm x 100mm 1000 to 4,500 RPM 140 to 2,800 xg

6-Place Fixed-Angle Rotor

7786047

2

75mm tube holder

7713043

1

6 17mm x 75mm 1000 to 4,000 RPM 140 to 1,500 xg

100mm tube holder

7713040

1

6 17mm x 100mm 1000 to 4,000 RPM 140 to 1,730 xg

125mm tube holder

7713044

2

6 17mm x 125mm 1000 to 4,000 RPM 140 to 1,950 xg

6-Place Performance Plus Horizontal 7786017

3

75mm tube holder

7713056

3

6 13mm x 75mm 1000 to 6,500 RPM 140 to 5,000 xg

12-Place combo Fixed-Angle STAT 7786003

3

75mm tube locations n/a

3

6 13mm x 75mm 1000 to 7,000 RPM 140 to 4,880 xg

100mm tube locations n/a

3

6 13mm x 100mm 1000 to 7,000 RPM 140 to 5,700 xg

12-Place Fixed-Angle STAT

7786001

3

75mm tube locations n/a

3

12 13mm x 75mm 1000 to 7,000 RPM 140 to 4,880 xg

Overall Dimensions (H x W x D):

8.0 in. x 11.0 in. x 13.5 in.

Centrifuge Motor:

1/2 H.P. Brushless DC

Protection Breaker:

4 Amp. re–settable

Timer:

electronic, with hold

or 0 to 99 minutes, +/- 1%

Weight:

23 lbs.

Permitted Environmental Conditions

Ambient Temperature During Operation: 2

o

C - 40

o

C

Maximum Relative Air Humidity:

90%

Electrical Requirements:

Power:

200 Watts

Voltage:

North America: 115 Volts (+/- 10%)

Japan:

100 Volts (+/- 10%)

Europe:

230 Volts (+/- 10%)

Frequency:

50/60 Hz

Any use other than those specified by the Manufacturer is explicitly prohibited.

Maximum sample density is 1.15 grams / mL, (water density = 1.0 grams / mL)

Specifications:

General Specifications for the Horizon Model 842VES Centrifuge

page 4

75 mm Tube

Holder

p/n 7713043

1

100 mm

Tube Holder

p/n 7713040

1

125 mm

Tube Holder

p/n 7713044

2

13 x 75mm

Performance Tube

Holder

p/n 7713056

3

Tube Holder

Cap

p/n 7713011

1” Cushion

Spacer

p/n 1525

1

This rotor or rotor accessory is rated for a rotational speed of 5000 RPM and a force of 3000 xg.

2

This rotor or rotor accessory is rated for a rotational speed of 4500 RPM and a force of 2500 xg.

3

This rotor or rotor accessory is rated for a rotational speed of 7000 RPM and a force of 5500 xg.

Refer to serial number label

on rear of centrifuge to

confirm voltage requirement.

NOTE: Contact the Drucker Company for assistance when changing the installed rotor. The installed rotor may

only be changed by qualified personnel.

.5 to 1 ml tube

Adapter

p/n 7713068

13 x 75mm

insert, p/n

7713064

13 x 100 mm

insert, p/n

7713066

1.5 to 2mL

tube Adapter

p/n 7713065